Setup Mode Menus
91
5.4.3.4. SETUP
RNGE
UNIT: SETTING THE REPORTING RANGE UNITS OF MEASURE
The T200 can display and report concentrations in ppb, ppm, ug/m
3
, mg/m
3
units.
Changing units affects all of the COM port values, and all of the display values for all
reporting ranges. To change the units of measure press
:
The following equations give
approximate conversions
between volume/volume units
and weight/volume units:
NO
ppb x 1.34 = µg/m
3
ppm x 1.34 = mg/m
3
NO2
ppb x 2.05 = µg/m
3
ppm x 2.05 = mg/m
3
Concentration field
displays all gases.
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST
TST> CAL
SETUP
SETUP X.X
CONC UNITS:PPB
PPB PPM UGM MGM
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
DIL
EXIT
Toggle these buttons to
select the units of measure
for the reporting ranges.
EXIT discards the new
setting.
ENTR accepts the
new setting.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Concentrations displayed in mg/m3 and ug/m3 use 0
C
mmHg for
Standard Temperature and Pressure (STP).
Consult your local regulations for the STP used by your agency.
(Example: US EPA uses 25
C as the reference temperature).
Once the Units of Measurement have been changed from volumetric (ppb or ppm)
to mass units (ug/m3 or mg/m3) the analyzer MUST be recalibrated, as the
―expected span values‖ previously in effect will no longer be valid.
Simply entering new expected span values without running the entire calibration
routine is not sufficient. This will also counteract any discrepancies between STP
definitions.
5.4.3.5. SETUP
RNGE
DIL: USING THE OPTIONAL DILUTION RATIO FEATURE
The dilution ratio feature is a software utility option designed for applications where the
sample gas is diluted before being analyzed by the T200. Typically this occurs in
continuous emission monitoring (CEM) applications where the quality of gas in a smoke
stack is being tested and the sampling method used to remove the gas from the stack
dilutes the gas. Once the degree of dilution is known, this feature allows the user to add
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...