Getting Started
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
72
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for both the NO
and NO
x
measurements.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
NOX STB= XXX.X PPB
NOX=XXX.X
< TST TST > CAL SETUP
M-P CAL
NOX STB= XXX.X PPB
NOX=XXX.X
<TST TST>
ZERO
CONC
EXIT
EXIT
at this point
returns to the
SAMPLE menu.
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for both the NO
and NO
x
measurements.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE RANGE=500.0 PPB NOX= XXXX
< TST
TST >
CAL SETUP
Set the Display to show
the
NOX STB
test
function.
This function calculates
the stability of the NO/NO
x
measurement
Toggle
TST>
button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until
NOX STB
falls below
0.5 ppB.
This may take several
minutes.
SAMPLE
NOX STB= XXX.X PPB
NOX=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
NOX STB= XXX.X PPB
NOX=X.XXX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
NOX STB= XXX.X PPB
NOX=X.XXX
<TST TST> ZERO
SPAN
CONC
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until
NOX STB
falls below
0.5 PPB.
This may take several
minutes.
SAMPLE
NOX STB= XXX.X PPB
NOX=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
NOX STB= XXX.X PPB
NOX=X.XXX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
NOX STB= XXX.X PPB
NOX=X.XXX
<TST TST> ENTR
CONC
EXIT
The
SPAN
button now
appears during the transition
from zero to span.
You may see both buttons.
If either the
ZERO
or
SPAN
buttons fail to appear see the
Troubleshooting Section of
this manual.
Analyzer continues to
cycle through NO
x
,
NO, and NO
2
measurements
throughout this
procedure.
The T200 Analyzer is now ready for operation.
Note
Once you have completed the above set-up procedures, please fill out the
Quality Questionnaire that was shipped with your unit and return it to Teledyne
API. This information is vital to our efforts in continuously improving our
service and our products. THANK YOU.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...