Data Acquisition System (DAS) and APICOM
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
160
To define the
REPORT PERIOD
, follow the instruction shown in Section 7.1.3 then
press:
Starting at the
EDIT CHANNEL MENU
SETUP X.X
0) CONC: ATIMER 5, 800
PREV NEXT
INS
DEL
EDIT
PRNT
EXIT
SETUP X.X
NAME: CONC
<SET
SET>
EDIT
EXIT
SETUP X.X
REPORT PERIOD:000:01:00
<SET
SET
>
EDIT
EXIT
Use the
PREV
and
NEXT
buttons to
scroll to the
DATA
CHANNEL
to be
edited.
SETUP X.X
REPORT PERIOD DAYS:0
0
0
0
ENTR EXIT
Toggle these buttons to
set the days between
reports (0 – 366).
SETUP X.X
REPORT PERIOD TIME:01:00
0
1
:0
0
ENTR EXIT
Press buttons to set amount of
time between reports, in hours
(HH) and/or minutes (MM)
(max: 23:59). 01:00 sets a
report to be made every hour.
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
Continue pressing
SET>
until ...
The
SAMPLE PERIOD
and
REPORT PERIOD
intervals are synchronized to the
beginning and end of the appropriate interval of the instruments internal clock.
If
SAMPLE PERIOD
is set for one minute the first reading would occur at the
beginning of the next full minute according to the instrument’s internal clock.
If the
REPORT PERIOD
is set for of one hour, the first report activity would occur at
the beginning of the next full hour according to the instrument’s internal clock.
EXAMPLE:
Given the above settings, if DAS parameters are activated at 7:57:35 the first sample
would occur at 7:58 and the first report would be calculated at 8:00 consisting of data
points for 7:58. 7:59 and 8:00.
During the next hour (from 8:01 to 9:00), the instrument will take a sample reading
every minute and include 60 sample readings.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...