4-45
Processing Instructions
Operator Manual
920011-649
alternating with:
and printer records:
==========================
CYCLE PAUSE
HH:MM:SS XM
==========================
2.
If it is safe to resume cycle
, display shows:
Press
START
touch pad to resume cycle. Printer records:
========================
CYCLE RESUME HH:MM:SS XM
========================
... and the cycle continues with the next phase in the sequence
until the cycle is completed.
...or...
3. Press
STOP
touch pad a second time to abort cycle.
Following Abort sequence, processor performs steps below
one
or two
times, depending on which cycle phase was in progress:
•
chamber is filled with filtered hot tap water
•
water is recirculated
•
water is drained
NOTE:
Procedure takes 7 to 15 minutes depending on cycle
phase in progress. Door remains locked until end of Abort
sequence(s).
4.
If cycle cannot be resumed
, display shows:
Press
STOP
touch pad to abort cycle.
During Abort sequence, display shows:
STOP PRESSED
CYCLE DELAYED
START TO RESUME
STOP TO CANCEL
STOP TO CANCEL
CANCELLING
CYCLE
WARNING – INEFFECTIVE
PROCESS HAZARD:
•
Correct
function
of
the
system is only ensured after
any diagnosed errors have
been found, corrected and a
successful processing cycle
has been completed.
•
Endoscopes and accessories
are not adequately processed,
rinsed and air purged when
processing cycle is aborted.
REPROCESS LOAD.
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...