Operator Manual
Troubleshooting
Alarm sounds, and display shows:
FAULT 30
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 30 =
0.2 µm FILTER FAILURE
1. Try to resume cycle by pressing
START
touch pad. If
situation reoccurs, abort cycle.
2. 0.2
μm
bacterial-retentive filter is defective. Replace filter
cartridge. Refer to
S
ECTION
for procedure.
3. If situation reoccurs, confirm that venting connection is
attached to filter housing cap and filter housing cap is
properly installed (refer to
S
ECTION
). If situation
reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 31
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 31 =
AIR PURGE RTD HEATING MALFUNCTION
Try to resume cycle by pressing
START
touch pad. If
situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 32
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 32 =
BOOT A: HIGH PRESSURE
Possible causes:
1. An endoscope of a diameter larger than template is
processed with insertion tube sleeve(s). Do not use
insertion tube sleeve for these endoscopes (refer to
S
ECTION
).
2. Relief valve is not working properly. Call STERIS.*
3. Relief valve pressure incorrectly adjusted. Call STERIS.*
4.
If 60 Hz processor
, ensure there is a spacer between
valve cap and valve stem.
5. Try to resume cycle by pressing
START
touch pad. If
situation reoccurs, call STERIS.*
*Service charges may be incurred. Consult your warranty for details.
FAULT
POSSIBLE CAUSE AND CORRECTION
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...