4-31
Processing Instructions
Operator Manual
920011-649
4. To process
one
endoscope in a cycle
with a washing phase
,
press
SINGLE
touch pad
once
. Display shows:
...or...
To process
one
endoscope in a cycle
without a washing phase
(high level disinfection phase only), press
SINGLE
touch pad
twice
. Display shows:
...or...
To process
two
endoscopes in a cycle
with a washing phase
,
press
DOUBLE
touch pad
once
. Display shows:
...or...
To process
two
endoscopes in a cycle
without a washing phase
(high level disinfection phase only), press
DOUBLE
touch pad
twice
. Display shows:
IMPORTANT: When WASHING SELECT option is set to NO in
Miscellaneous Values menu of Supervisor mode (refer to
), operator cannot
select an endoscope processing cycle WITHOUT WASHING
because this option has been disabled. Only SINGLE endoscope
processing cycle with washing phase and DOUBLE endoscope
processing cycle with washing phase are available. Pressing
touch pad a second time will have no effect.
NOTE:
For a 50 Hz processor, refer to Operator Manual
Information Supplement (P920501-322) for specific information.
5. If bar code reader accessory is not present, or is present but not
activated, go to
Step 8
.
If bar code reader accessory is present and activated, display
shows:
SINGLE
WITH WASHING
SINGLE
WITHOUT WASHING
DOUBLE
WITH WASHING
DOUBLE
WITHOUT WASHING
SCAN
OPERATOR ID
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...