5-5
Cycle and Control Value Programming
Operator Manual
920011-649
3. Press
ENTER
touch pad to confirm selection. Display moves on
to WASH segment of selected cycle.
Display shows, for example:
__ blinking
and printer records, for example:
==========================
HH:MM:SS
XM MM-DD-YY
==========================
MODIFY: SINGLE
==========================
4. Press
SCROLL arrows
(left or right) to change time value of
WASH 1 segment (adjustable from five to ten minutes).
NOTE: If a value is out of range, display shows OUT OF
RANGE–SEE OPERATOR MANUAL
for four seconds, then
control returns to current display.
5. Press
ENTER
touch pad to confirm new WASH 1 segment time
value. Printer records:
============================
CYCLE MODIFICATION SAVED
============================
NOTE: For new values to be saved, it is necessary to scroll through
all phases of a cycle and confirm values by pressing ENTER touch
pad after each phase, even if no value was changed.
- or -
To cancel a modification, press
ESC
touch pad once. Display
returns to cycle being modified. Values are not saved. Printer
records:
============================
CYCLE MODIFICATION NOT
SAVED
============================
6. Display moves on to AIR PURGE phase time of selected cycle.
Display shows:
__ blinking
WASH
RECIRC MM:SS
AIR PURGE
RECIRC 04:00
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...