5-7
Cycle and Control Value Programming
Operator Manual
920011-649
If
YES
is selected, printer records values of cycle just modified,
for example:
===========================
HH:MM:SS XM
MM-DD-YY
===========================
CYCLE - SINGLE
WITH WASHING
===========================
WASH
WATER TYPE
HOT
TEMPERATURE
48.0 C
RECIRC. TIME
05:00
CHEMICAL
KLENZYME
CONCENTR.
2 mL/L
RINSE
WATER TYPE
HOT
RECIRC. TIME
00:40
HLD
WATER TYPE
HOT
TEMPERATURE
50.0 C
GENERATION
04:00
DISINFECTION
06:00
RINSE 1
WATER TYPE
HOT
RECIRC. TIME
00:40
RINSE 2
WATER TYPE
HOT
RECIRC. TIME
00:40
AIR PURGE
RECIRC. TIME
04:00
===========================
Display shows:
__ blinking
12. Press
SCROLL arrows
(up or down) to choose another
Supervisor mode menu.
- or -
Press
ESC
touch pad to exit Supervisor mode.
SCROLL TO MODIFY
SINGLE