5-4
920011-649
Operator Manual
Cycle and Control Value Programming
then:
then:
3. Press
CHANGE VALUES
touch pad to enter Supervisor mode.
Display shows for example:
__ blinking
and printer records:
===========================
HH:MM:SS
XM MM-DD-YY
===========================
= SUPERVISOR MODE =
===========================
5.2.2 Changing Cycle Values
Changes affect only the selected cycle. Repeat procedure to
change values of other cycles.
Change Cycle Values displays are used for following functions (see
Cycle Description Chart
,
for available value ranges):
1. Access Supervisor mode as explained in
. Display shows:
__ blinking
2. Press
SCROLL arrows
(up or down) to choose SINGLE or
DOUBLE.
NOTE: It is also possible to select cycle by pressing directly on
corresponding cycle touch pad.
STERIS CORP
COPYRIGHT (C) 2011
READY
MM-DD-YY
HH:MM XM
SCROLL TO MODIFY
SINGLE
DISPLAY
FUNCTION
Recirculation Time
To set recirculation time for WASHING
segments WASH 1 and 2 (from 5 to
10 minutes) and AIR PURGE (from 4 to
30 minutes).
Print Cycle
To print values of currently modified cycle.
SCROLL TO MODIFY
SINGLE
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...