4-25
Processing Instructions
Operator Manual
920011-649
4.5 Installing Reliance
Dry Chemistry
Container
Disinfection solution is generated within the Reliance EPS.
IMPORTANT: Read Reliance dry chemistry package insert for
detailed instructions for use.
Following directions for use in the Reliance dry chemistry package
insert, prepare and insert container into the processor Reliance CDS.
NOTE:
Before inserting Reliance dry chemistry container, verify the
three punching rods inside the Reliance CDS are tight. If not, refer
to
, for procedure.
Once the Reliance dry chemistry container is inserted, close and
latch Reliance CDS.
WARNING – INEFFECTIVE
PROCESS HAZARD:
•
Only
use
Reliance
dry
chemistry in Reliance EPS
Processor. Do not substitute
any
other
product
for
Reliance dry chemistry.
•
Inspect
Reliance
dry
chemistry
container
for
damage prior to use. Do
not
use
damaged
containers; if damage is
observed, follow directions
for disposal in Reliance dry
chemistry package insert.
•
Storage of Reliance dry
chemistry container should
be in a clean dry location.
Reliance
dry
chemistry
container should not be
exposed to extreme heat or
humidity. Refer to package
insert for instructions.
•
Before starting a cycle,
ensure to verify Reliance
dry chemistry container's
expiration date and STERIS
chemical indicator bottle's
expiration date are valid. If
expiration date is not valid,
use a new Reliance dry
chemistry container or a
new
chemical
indicator
strip. Failure to do so may
result
in
ineffective
processing.
WARNING
–
CHEMICAL BURN
AND/OR EYE INJURY HAZARD:
All personnel involved in the
operation of this processor
should be able to demonstrate
the
correct
procedure for
disposal
of
Reliance
dry
chemistry containers.
Figure 4-22. Reliance Dry Chemistry
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...