4-7
Processing Instructions
Operator Manual
920011-649
10. Determine if endoscope requires an insertion tube sleeve as
follows (see Figures 4-4 and 4-5):
NOTE: Insertion tube should fit easily with no force applied.
•
Locate three diameter template slots in front of endoscope
processing support.
•
Place portion of insertion tube closest to bending rubber into
smaller of template slots. If insertion tube fits in this slot, use
insertion tube sleeve labeled with a
RED
dot.
•
If insertion tube does not fit into this slot, go to next template slot.
•
If insertion tube fits into next template slot, use insertion tube
sleeve labeled with a
BLUE
dot.
•
If insertion tube does not fit into this slot, go to next template slot.
•
If insertion tube fits in this larger slot, use insertion tube
sleeve
NOT labeled
with a color dot
(split insertion tube
sleeve).
NOTE: Some endoscopes have a distal tip that is larger than
the insertion tube diameter. These endoscopes will fit into a
template slot but the distal tips will not fit into the designated
insertion tube sleeve. Use the split insertion tube sleeve
(NOT labeled with a color dot) to process these endoscopes.
•
If insertion tube does not fit into any template slot, endoscope
does not need an insertion tube sleeve.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE: When using insertion
tube sleeve, ensure that sleeve
handle
is
pushed
down.
Insertion
tube
must
be
inserted over sleeve handle to
avoid possible damage.
Insertion Tube
Distal Tip
Insertion Tube
Figure 4-4. Carefully Insert Insertion Tube in Template Slot
Front o
f Endo
scope
Proce
ssing S
upport
Insertion Tube
Diameter
Template Slots
Place Insertion
Tube (NOT
Insertion Tube
Distal Tip) in
Template Slot
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...