4-34
920011-649
Operator Manual
Processing Instructions
4.8.4 Low Chemical Warning
This option is used to warn operator when level of chemical in Bag-
in-a-Box indicated in display message is getting low. It should
prevent having a chemical empty alarm during a cycle.
Low chemical warning is enabled in
Miscellaneous Values
menu of
Supervisor mode by setting
CHEMICAL WARNING option
to
YES
(refer to
).
Level of chemical in Klenzyme
®
Enzymatic Presoak and Cleaner
Bag-in-a-Box is automatically verified during start-up of SINGLE and
DOUBLE cycles with WASHING PHASE. Level of chemical in CIP
200
®
Acid-Based Process and Research Cleaner Bag-in-a-Box is
automatically verified during start-up of D-LONG cycle.
1. If chemical level is low when
START
touch pad is pressed,
display shows:
or:
alternating with:
and:
IMPORTANT: When previous messages are displayed,
approximately five cycles can be processed before indicated
Bag-in-a-Box is completely empty. "FAULT 08 or 09, CYCLE
DELAYED" alarm is triggered, indicating a Bag-in-a-Box is empty.
2. Replace indicated Bag-in-a-Box as explained in
.
or:
Press
STOP
touch pad to ignore message and process selected
cycle.
KLENZYME
IS LOW
CIP 200
IS LOW
REPLACE AND
PRESS START
PRESS STOP
TO IGNORE
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...