4-24
920011-649
Operator Manual
Processing Instructions
5. In Idle mode (when READY is shown on display), activate leak
test self-diagnostic by pressing
START
touch pad four times. A
message is displayed to indicate self-diagnostic is running:
The following printout is printed at the end of the leak test self-
diagnostic:
========================
HH:MM:SS XM MM-DD-YY
========================
= LEAK TEST =
= SELF-DIAGNOSTIC =
========================
========================
TEST 1: XXXX
TEST 2: XXXX
========================
XXXX shows "PASS" for passing tests or "FAIL" for failed tests.
Repeat
to perform leak test self-diagnostic with hose
assembly using second hose assembly.
LEAK TEST
SELF-DIAGNOSTIC