4-23
Processing Instructions
Operator Manual
920011-649
4.4.6 Self-Diagnostic
The leak test self-diagnostic is available to the user and is to be
performed without any endoscope to verify the integrity of the
system. The leak test self-diagnostic lasts approximately 90
seconds. Leak test self-diagnostic can be performed with either the
door opened or closed.
Performing such a test without any hose assembly connected in the
processor allows the verification of the system.
To proceed with leak test self-diagnostic without any hose assembly
connected, go directly to
Step 5
Running the same test with the hose assembly connected and using
the plug provided on the other extremity allows verification of the
hose assembly. See
for details on how
to connect hose assembly and see
.
To run leak test self-diagnostic using a hose plug on hose assembly,
proceed as follows:
1. Select appropriate leak test hose (see
and Figure 4-16 for details).
2. Connect quick connect extremity of leak test hose(s) to flexible
connection(s) labeled 1 and/or 2 (see
for details).
3. Select appropriate hose plug.
For example, hose plugs are identified "O" for Olympus
®
endoscopes, "P" for Pentax
®
endoscopes or "F" for Fujifilm
®
endoscopes(see Figure 4-21).
4. Connect hose plug to leak test hose other extremity.
NOTE:
Hose plug must remain in place on the extremity of the
leak test hose for the duration of the leak test self-diagnostic.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE:
Remove
any
automated leak test hoses
present
in
endoscope
processing support before
performing a D-SHORT or a D-
LONG cycle. Performing a
decontamination
cycle
(D-
SHORT and D-LONG) with
automated leak test hose
assemblies
present
may
cause water to enter hoses.
Inspect
leak
test
hose
assemblies for presence of
water prior to processing a
cycle. Water in tubing may
cause damage to endoscopes.
REF: 920-502-707
Figure 4-21. Automated Leak Test Hose Assembly
Plug (Olympus Shown)
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...