Processing Instructions
Following the loading in the endoscope processing support, connect
leak test hoses 1 and/or 2 to endoscope(s) as follows:
1. Identify endoscope(s) manufacturer.
2. Select appropriate leak test hose(s).
For example, leak test hoses are identified "O" for use with
Olympus
®
endoscopes, "P" for use with Pentax
®
endoscopes or
"F" for use with Fujifilm
®
endoscopes (see Figure 4-16).
3. Slightly shake chamber flexible connections which are intended
to be used so any water droplets, remaining from preceding
cycle, do not get trapped when connecting leak test hoses.
4. To connect quick connect extremity of leak test hose(s) to
flexible connection(s) labeled 1 and/or 2, proceed as follows:
a. With one hand, hold quick connect extremity of leak test
hose (see Figure 4-17).
b. With other hand, press flexible connection into quick
connect. Quick connect locking sleeve moves forward and
locks automatically (see Figure 4-17).
IMPORTANT: Only endoscopes with a leak test port on the
light guide cable can be leak tested.
5. Connect other extremity of leak test hose(s) to endoscope leak
test port(s) (see Figure 4-18).
6. Insert Reliance dry chemistry container and STERIS
chemical indicator strip in endoscope processing support
(see
,
and
for details).
CAUTION – POSSIBLE EQUIPMENT
DAMAGE:
Remove
any
automated leak test hoses
present
in
endoscope
processing
support
before
performing a D-SHORT or a D-
LONG cycle. Performing a
decontamination
cycle
(D-
SHORT and D-LONG) with
automated
leak
test
hose
assemblies present may cause
water to enter hoses. Inspect
leak test hose assemblies for
presence of water prior to
processing a cycle. Water in
tubing may cause damage to
endoscopes.
Figure 4-16. Automated Leak Test Hose Assembly
(Fujifilm Shown)
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...