Operator Manual
Troubleshooting
FAULT 53 Continued...
LEAK TEST 2 FAIL (Cont’d)
5. Perform self-diagnostic to determine if leak comes from
hose assembly (use plugs already provided). If yes,
change and lubricate endoscope connection O-ring (refer
to
S
ECTIONS
and
) or change hose assembly.
6. Ensure hose plug remains properly in place on extremity of
leak test hose for all duration of self-diagnostic test. If
unsuccessful, go to
if necessary.
Alarm sounds, and display shows:
FAULT 54
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 54 =
LEAK TEST PORT 1: HOSE ASSEMBLY DETECTED
Press
ALARM REPLY
touch pad to silence audible alarm.
Possible causes:
1. Endoscope present in processor (with hose connected)
during initiation of D-SHORT or D-LONG cycle.
2. Hose assembly 1 connected to processor.
Cycle is automatically aborted. User is required to:
1. Verify no endoscope is present in processor (with hose
connected). If yes, remove endoscope and hose assembly
and restart decontamination cycle. If not, go to
Step 2
2. Verify no hose assembly is connected to processor leak test
port 1. If yes, remove hose assembly, if not, go to
Step 3
3. Perform self-diagnostic without hose assembly to
determine if leak comes from processor system. If yes, call
STERIS.*
Alarm sounds, and display shows:
FAULT 55
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 55 =
LEAK TEST PORT 2: HOSE ASSEMBLY DETECTED
Press
ALARM REPLY
touch pad to silence audible alarm.
Possible causes:
1. Endoscope present in processor (with hose connected)
during initiation of D-SHORT or D-LONG cycle.
2. Hose assembly 2 connected to processor.
Cycle is automatically aborted. User is required to:
1. Verify no endoscope is present in processor (with hose
connected). If yes, remove endoscope and hose assembly
and restart decontamination cycle. If not, go to
Step 2
2. Verify no hose assembly is connected to processor leak test
port 2. If yes, remove hose assembly, if not, go to
Step 3
.
3. Perform self-diagnostic without hose assembly to
determine if leak comes from processor system. If yes, call
STERIS.*
*Service charges may be incurred. Consult your warranty for details.
FAULT
POSSIBLE CAUSE AND CORRECTION
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...