Operator Manual
Processing Instructions
4.17 Disposing of
Expired Dry Chemistry
To dispose of expired Reliance dry chemistry,
run a DOUBLE
endoscope processing cycle
without
washing phase, and
without
endoscopes
.
See
,
through
, for additional information.
NOTE:
For a 50 Hz processor, refer to the Operator Manual Information
Supplement (P920501-322) for additional information.
IMPORTANT: For this special DOUBLE endoscope processing cycle,
test cords and insertion tube sleeves with a blue dot identification must
be used in both control handle boots "A" and "B". This will maintain
sufficient pressure inside control handle boots in absence of endoscopes.
To run a DOUBLE endoscope processing cycle
without
washing
phase, and
without endoscopes
, proceed as follows:
1. Verify display shows:
2. If chamber door is not open, press
UNLOCK
toggle switch to
unlock processor door and open door. Printer prints:
==========================
DOOR OPEN
HH:MM:SS XM
==========================
3. Pull endoscope processing support out of chamber.
4. Open control handle boots "A" and "B". Insert one insertion tube
sleeve
with a blue dot identification
in insertion tube inlet of
each boot (see Figure 4-27).
READY
MM-DD-YY
HH:MM XM
WARNING
–
CHEMICAL BURN
AND/OR EYE INJURY HAZARD:
All personnel involved in the
operation of this processor
should be able to demonstrate
the
correct
procedure for
disposal
of
Reliance
dry
chemistry containers.
Flaps
Insertion Tube Sleeve
With a Blue Dot
Identification
Figure 4-27. Insertion Tube Sleeve
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...