7-21
Troubleshooting
Operator Manual
920011-649
FAULT 48 Continued...
LEAK TEST PORT 1: PRESSURE NOT REACHED (Cont’d)
6. Perform self-diagnostic to determine if leak comes from
hose assembly (use plugs already provided). If yes,
change and lubricate endoscope connection O-ring (refer
to
S
ECTIONS
and
) or change hose assembly. If
not, call STERIS.*
7. Ensure hose plug remains properly in place on extremity of
leak test hose for all duration of self-diagnostic test. If
unsuccessful, go to
Step 5
.
Alarm sounds, and display shows:
FAULT 49
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 49 =
LEAK TEST PORT 2: PRESSURE NOT REACHED
Press
ALARM REPLY
touch pad to silence audible alarm.
Possible causes:
1. Hose assembly 2 not connected to endoscope.
2. Major leak in endoscope 2 or in system.
3. Air blockage within pneumatic block.
4. Hose plug not used or removed from extremity of leak test
hose assembly during self-diagnostic test. Go to
Step 7
Cycle is automatically aborted. User is required to:
1. Connect endoscope if not connected and restart
processing cycle. If unsuccessful, go to
Step 2
.
2. Verify soaking cap of endoscope. Replace soaking cap if
necessary and restart a processing cycle. If unsuccessful,
go to
Step 3
.
3. Perform dry leak test according to endoscope
manufacturer recommendations. If no leak is detected, go
to
Step 4
4. Perform wet leak test according to manufacturers
recommendations. If no leak is detected, go to
Step 5
.
5. Perform self-diagnostic without hose assembly to
determine if leak comes from processor system. If yes, call
STERIS.* If not, go to
Step 6
.
6. Perform self-diagnostic to determine if leak comes from
hose assembly (use plugs already provided). If yes,
change and lubricate endoscope connection O-ring (refer
to
S
ECTIONS
and
) or change hose assembly. If
not, call STERIS.*
7. Ensure hose plug remains properly in place on extremity of
leak test hose for duration of self-diagnostic test. If
unsuccessful, go to
Step 5
.
*Service charges may be incurred. Consult your warranty for details.
FAULT
POSSIBLE CAUSE AND CORRECTION
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...