7-22
920011-649
Operator Manual
Troubleshooting
Alarm sounds, and display shows:
FAULT 50
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 50 =
LEAK TEST PORT 1: EXHAUST TOO LONG
Press
ALARM REPLY
touch pad to silence audible alarm.
Possible causes:
1. Exhaust valve malfunction (leak test port 1).
2. Air blockage within pneumatic block.
Cycle aborted; users are required to restart a processing
cycle. If unsuccessful, call STERIS.*
Alarm sounds, and display shows:
FAULT 51
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 51 =
LEAK TEST PORT 2: EXHAUST TOO LONG
Press
ALARM REPLY
touch pad to silence audible alarm.
Possible causes:
1. Exhaust valve malfunction (leak test port 2).
2. Air blockage within pneumatic block.
Cycle aborted; users are required to restart a processing
cycle. If unsuccessful, call STERIS.*
Alarm sounds, and display shows:
FAULT 52
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 52 =
LEAK TEST 1 FAIL
Press
ALARM REPLY
touch pad to silence audible alarm.
Pressure loss over acceptance criterion during monitoring
time.
Possible causes:
1. Endoscope leaks.
2. Hose plug removed from extremity of leak test hose
assembly during self-diagnostic test. Go to
Step 6
Continued...
*Service charges may be incurred. Consult your warranty for details.
FAULT
POSSIBLE CAUSE AND CORRECTION
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...