1-5
Safety Precautions
Operator Manual
920011-649
CAUTION – POSSIBLE EQUIPMENT DAMAGE (Cont’d):
Use silicone lubricant to lubricate squeeze tubes. Petroleum-based lubricants, such as petroleum jelly
or grease, will damage squeeze tubes.
Wear clean gloves when replacing air supply filter cartridge. Do not touch cartridge with bare hands to
prevent oils from being deposited.
When installing or replacing 0.2 µm bacterial-retentive filter cartridge, do not touch cartridge with bare
hands to prevent oils from being deposited. Open plastic wrapping and hold filter cartridge through
wrapping when inserting into filter housing.
When using insertion tube sleeve, ensure that sleeve handle is pushed down. Insertion tube must be
inserted over sleeve handle to avoid possible damage.