4-37
Processing Instructions
Operator Manual
920011-649
CYCLE – SINGLE:
Name of endoscope
processing cycle (SINGLE cycle).
WITH WASHING:
Name of phase.
Includes WASH and RINSE segments.
CYCLE START:
Time cycle was started.
CYCLE DATE:
Date cycle was started.
CYCLE NUMBER:
Unique sequential
number starting at 00001.
UNIT NUMBER:
Serial number of
processor.
If bar code reader accessory is
present and activated:
OPERATOR ID:
Operator's code
number and name.
BOOT A:
Information printed below is
for endoscope in control handle boot
"A".
PATIENT ID:
Patient name.
DEVICE ID:
Identification number of
endoscope.
PROCEDURE
ID:
Identification
number of procedure.
PHYSICIAN ID
: Physician's code and
name.
If bar code reader accessory is
present and activated, and DOUBLE
cycle is selected, information is also
printed for control handle boot "B".
PHASE:
Name of phase and segment
processed with time it started:
•
WASH =
Washing with hot filtered
water and detergent.
•
RINSE =
Washing Rinse with hot
filtered water.
If DOUBLE WASHING feature is
activated, this sequence is repeated
twice (refer to
S
ECTION
5.2.3, S
ETTING
M
ISCELLANEOUS
V
ALUES
).
•
HLD =
HIGH-LEVEL DISINFECTION
phase of SINGLE/DOUBLE cycle.
•
RINSE 1 =
RINSE 1 with hot filtered
water.
•
RINSE 2 =
RINSE 2 with hot filtered
water.
•
AIR PURGE =
Air purge phase.
WATER TYPE:
Type of water used in
phase.
TEMPERATURE:
Temperature set
point during heated phases.
CHEMICAL:
Name of chemical
additive used for wash segment.
CONCENT.:
Concentration
of
chemical additive injected for wash
segment.
GENERATION:
Duration of high level
disinfection solution generation.
DISINFECTION:
Duration of exposure
to high level disinfection solution.
RECIRC. TIME:
Duration of solution/air
recirculation.
COMPLETED:
Time at which cycle
was completed.
TOTAL CYCLE:
Total duration of
endoscope processing cycle.
MAXIMUM FILLING TIME:
Maximum
filling time of processor during cycle.
Used as an indicator for degree of
clogging of dual pre-filter and
pressure
regulator
assembly
cartridges.
STOP PRESSED:
Printed if operator
pressed
STOP
touch pad to interrupt
air purge before maximum 60-minute
duration has elapsed.
SAFETY DOOR LOCK DISENGAGED:
Safety door lock is disengaged after
four minutes of air purge. Operator can
open door by pressing
UNLOCK
toggle switch at this point.
If EXTENDED PRINTOUT feature was
enabled,
additional space is printed to
allow operator to write down additional
information (titles and codes indicating
device[s] status are also added):
Codes are as follows:
OT =
Outdated device
AH =
After hours device
W =
Weekend device
PP =
Device processed prior to
procedure
RR =
Device returned from repair.
In question
DID CI PASS YES/NO;
CI =
STERIS chemical indicator
READY
TO
UNLOAD:
Operator
message to indicate load can be taken
out of chamber.
DOOR OPEN:
Time at which operator
opened chamber door.
Figure 4-25
. Typical Cycle Printout – Details
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...