6-15
Routine Maintenance
Operator Manual
920011-649
6.10.2 0.2 µm Bacterial-
Retentive Filter Cartridge
See Figure 6-7.
0.2 μm bacterial-retentive filter must be replaced at least every
90 days (default, adjustable from 0 to 90 days in Supervisor mode;
refer to
). This time
period may be shorter depending on water quality. A filter integrity
system also triggers a maintenance due message when cartridge
needs to be changed (refer to
). To
change cartridge, proceed as follows:
1. Wear appropriate Personal Protective Equipment (PPE) when
replacing 0.2 μm bacterial-retentive filter cartridge.
2. If processor is
ON
, press bottom portion of
POWER (ON/OFF)
toggle switch to turn processor
OFF
. Put processor main
disconnect switch to
OFF
position.
3. Remove screw holding service access panel to frame and
remove panel.
4. Disconnect venting connection of filter integrity system from
housing cap (see Figure 6-7).
5. Turn filter housing cap counterclockwise to loosen. Remove cap
(see Figure 6-7).
6. Pull out old filter cartridge and discard.
7. Open plastic bag containing new filter cartridge from end
closest to filter cartridge O-ring.
8. With filter still in plastic bag, slightly wet O-ring with water to ease
insertion.
9. Hold cartridge through bag. Insert new filter cartridge, O-ring
first, into filter housing. Turn 1/4-turn clockwise while pushing.
Remove plastic bag and discard.
10. To reinstall filter housing cap, turn clockwise until fully
tightened.
11. Reconnect venting connection to filter housing cap.
12. Ensure to wipe up any spilled water around filter housing and
floor area.
13. Reinstall service access panel with previously removed screw.
14. Put processor main disconnect switch to
ON
position. Press top
portion of
POWER (ON/OFF)
toggle switch to turn processor
ON
.
15.
Reset 0.2 μm bacterial-retentive filter cycle counter in Supervisor
mode (refer to
).
WARNING
–
PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD: When working with
compressed air, flying debris
and particles may present a
hazard
to
personnel
in
immediate area. Personal eye
protection must be worn when
working with compressed air.
WARNING
–
ELECTRIC SHOCK
AND/OR BURN HAZARD:
•
Fasteners and star washers
are used to ensure protective
bonding continuity. Always
reinstall any star washer
which may
have
been
removed during installation
or servicing.
•
Pressing bottom portion of
POWER (ON/OFF) toggle
switch to turn off processor
DOES NOT cut off electrical
power.
Lockout/Tagout
building electrical disconnect
switch to turn off power to
processor for maintenance
or repairs.
CAUTION
–
POSSIBLE EQUIPMENT
DAMAGE:
•
When installing or replacing
0.2 µm
bacterial-retentive
filter cartridge, do not
touch cartridge with bare
hands to prevent oils from
being
deposited.
Open
plastic wrapping and hold
filter
cartridge
through
wrapping when inserting
into filter housing.
• Before removing 0.2 µm
bacterial-retentive
filter
housing
cap,
always
disconnect air vent for filter
integrity system.
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...