6-5
Routine Maintenance
Operator Manual
920011-649
7. Close chamber door.
8. Press
DECONTAM
touch pad once. Display shows:
IMPORTANT: If a D-SHORT cycle has not been completed within
54 hours, processor control will be automatically locked, i.e. no
SINGLE or DOUBLE endoscope processing cycle can be started
until a D-LONG cycle is performed (see S
).
9. Press
START
touch pad. Display shows:
alternating with:
10. Confirm endoscopes, accessories and Flow Units are not
present within endoscope processing support.
11. Press
START
touch pad to start D-SHORT cycle. Processor
automatically proceeds through programmed phases of
D-SHORT cycle (see Figure 6-1 for sample printout). After air
purge phase is complete, buzzer sounds and display shows:
then:
Processor cools for 30 seconds. Then display shows:
12. Processor drains for 20 seconds. After draining phase is
completed, buzzer sounds and display shows:
13. Press
UNLOCK
toggle switch to unlock processor door. Open door.
14. Check debris screen for debris. Rinse screen under running
water if necessary. Rinse screen while it is still wet, before debris
dries. Return screen after cleaning.
D-SHORT
PRESS START
REMOVE ALL
ENDOSCOPES
PRESS START
WHEN READY
CYCLE COMPLETE
PRESS STOP
COOLING
PLEASE WAIT
DRAINING
PLEASE WAIT
CYCLE COMPLETE
READY TO UNLOAD
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...