7-17
Troubleshooting
Operator Manual
920011-649
Alarm sounds, and display shows:
FAULT 37
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 37 =
PROTECTION RTD DEFECTIVE
Try to resume cycle by pressing
START
touch pad. If
situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 38
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 38 =
FRESH DRY CHEMISTRY CONTAINER NOT DETECTED
A new Reliance dry chemistry container was not inserted in
Reliance CDS prior to cycle start. Cycle is aborted.
Start new cycle after ensuring a new Reliance dry
chemistry container is installed in Reliance CDS and
ensuring latch is closed on Reliance CDS. If situation
reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 39
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 39 =
DRY
CHEMISTRY
CONTAINER
DETECTION
DEFECTIVE
Try to resume cycle by pressing
START
touch pad. If
situation reoccurs, call STERIS.*
*Service charges may be incurred. Consult your warranty for details.
FAULT
POSSIBLE CAUSE AND CORRECTION
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...