4-27
Processing Instructions
Operator Manual
920011-649
4.7 Pre-Processing
Checklist
The following conditions are necessary to achieve successful
processing of the endoscopes and accessory types listed in the
Endoscope Quick Reference Guide
.
– ENDOSCOPE PREPARATION –
Endoscope and accessories were:
�
inspected and found to be in proper working order;
�
prepared for immersion;
�
successfully leak-tested;
�
manually pre-cleaned within 30 minutes prior to starting
processing cycle.
– PRIOR TO PROCESSING –
�
Debris screen
is clear of debris (see
);
�
Spray arms
rotate freely (see
);
�
Accessories
are correctly positioned in
endoscope processing
support
(see
);
�
Insertion tube sleeve
(labeled with a RED dot, labeled with a
BLUE dot or NOT labeled)
is used for endoscopes which can
be inserted into one of three template slots located in front of
endoscope processing support. If required, place proper
insertion tube sleeve into control handle boot "A" and/or "B" (see
);
�
Endoscope #1
is placed in control handle boot "A" (see
);
�
If
endoscope #1 does not have a light guide cable
, a
control handle boot plug is placed in the universal cord port;
�
Endoscope #2
is placed in control handle boot "B" (see
);
�
If
endoscope #2 does not have a light guide cable
, a
control handle boot plug is placed in the universal cord port;
�
If
only one endoscope is processed
, the endoscope is placed
in control handle boot "A" and a control handle boot plug is
placed in the fluid delivery port of control handle boot "B";
�
If
only accessories, such as dilators and bougies, are being
processed (without an endoscope)
, test cords are placed in
both control handle boots (follow instructions in
Endoscope
Quick Reference Guide
for Flow Unit 5)
;
�
If required,
Flow Units
are attached (refer to
Endoscope Quick
Reference Guide
);
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...