79
Table 47. Isolation Data
Isolation
Insulation Type
Maximum Working Voltage [V]
E1,E2,E3
1 MOPP
240 VAC + 5 VDC
E1
2 MOPP
5 VDC
F
2 MOOP
5 VDC
MOP = Means of Protection
MOOP = Means Of Operator Protection
MOPP = Means Of Patient Protection
File Formats
Table 48. Default spreadsheet file format
Text format:
ASCII
Data Rate:
100Hz
Separation:
TAB
Data Columns:
8 (Time, Pa, Pa mean, Pd, Pd mean, Pd/Pa, Heart rate, ECG)
Table 49. Units
Time:
seconds (decimal point)
Pa:
mmHg (decimal point)
Pa mean:
mmHg (decimal point)
Pd:
mmHg (decimal point)
Pd mean:
mmHg (decimal point)
Pd/Pa:
N/A (decimal point)
Heart rate:
bpm
ECG:
mV
DICOM
See DICOM Conformance Statement at sjmprofessional.com
Compliance with Regulatory Requirements
Compliance with Standards and Directives
St. Jude Medical hereby declares that the QUANTIEN™ measurement system is in compliance
with the essential requirements and other relevant provisions of the Medical Device Directive (EC
Directive 93/42/EEC), RE directive 2014/53/EU Annex II, and Quality System Regulation (US). A
copy of the Full Declaration of Conformity can be obtained by contacting the EU Representative.
QUANTIEN measurement system electrical safety classifications:
Class II equipment; the type of protection against electric shock.
State: Released
Date: 2019.03.06 22:30 GMT