Isolation: IEC 60601-1 Type CF Defibrillation protected
AUX IN 1 (AO)
Differential analog input
Scale factor: Selectable between 50 mmHg/V and 200 mmHg/V
Electrical range: ± 10 V (Depending on selected scale factor the measurement range can be less)
Isolation: IEC 60601-1 Basic isolation
AUX IN 2 (ECG)
Differential analog input
Scale factor: 1 mV / V
Electrical range: ± 10 V
Isolation: IEC 60601-1 Basic isolation
AUX OUT 1 & 2
Analog output
Signal source: Selectable Pa or Pd.
Scale factor: 1 V/100 mmHg
Max load: 10 kOhm/100 nF
Isolation: IEC 60601-1 Basic isolation
USB
Version: 2.0, supports Hi-speed communication
Connector: Type A, Receptacle
Ethernet
10/100 Mbit/s
Connector: RJ45
DVI
Screen resolution of QUANTIEN™ main unit: 1280x800
Output resolutions (DVI out): Auto
Frequency: 60 Hz
Connector: DVI-I (Integrated - digital and analog)
Connect with a standard VGA cable and DVI to VGA adapter (analog signal)
Connect with a DVI to VGA cable (analog signal)
Connect with a DVI-D cable (digital signal)
Connect with a DVI to HDMI cable (digital signal)
Mains Connector in Power Supply Unit
Connector: Standard inlet IEC 320-C8
Wireless Transmission
Frequency range: 2.4000-2.4835 GHz (ISM-band)
Type: Frequency Hopping Spread Spectrum
PressureWire™ wireless guidewire and Wi-Box™ radio transmitter specification see IFU for
PressureWire wireless guidewire and Wi-Box™ AO transmitter.
Mounting
VESA 75x75 mm and 100x100 mm
State: Released
Date: 2019.03.06 22:30 GMT