The slow sweep is used during RFR pullback regardless of user setting.
Options:
Slow (40 seconds/screen)
Medium (20 seconds/screen)
Fast (10 seconds/screen)
ECG (AUX IN 2)
ECG Configuration
Scale
. Select the desired ECG scale.
Select the
Display Trace
checkbox to display ECG signal in Live Window.
Select the
Baseline Filter
checkbox to enable high pass filter for baseline drift removal.
Pressure
Average
: Number of heart beats included in pressure averaging filter.
Select 1, 2, 3, 4, 8 or 12 beats. Factory setting: 3 heart beats.
Choosing a high number of heartbeats will make the pressure averaging slower and less
sensitive to artifacts, but may also result in insensitive averaging which is noticeable when
there is a short hyperemic plateau.
Choosing a low number of heartbeats makes the pressure averaging faster and more
sensitive to pressure changes, which is desirable when using a short hyperemic plateau, but
it may also result in an averaging overly sensitive to arrhythmia and pressure disturbances.
CAUTION: An insensitive or overly sensitive averaging of pressure may result in an
incorrect FFR value.
Scale:
Pressure waveform vertical scale.
Select -20-80; 0-100; 0-150; 0-200; 0-200, -80-320 mmHg. Factory setting: 0-150 mmHg.
Display Max/Min in Review
: Enable indicators for minimum and maximum
diastolic/systolic pressure indicators in Review mode.
Trend Graph
Display Trend Graph
: Turn on/off the lower Trend Graph window
Pd/Pa Scale
: Change the scale of the Pd/Pa graph
Brightness
Increase or decrease screen brightness.
Select 20, 40, 60, 80 or 100%. Factory setting is 60%.
Other Settings
FFR Set Up Guide:
Turn on/off FFR procedure step-by-step guide
Hide Patient Information:
Hide patient information on screen. Useful during live case
broadcasting.
Max Rec Length
(settings in
(Setup)
Menu,
System, Other
):
-
Maximum recording time. If a recording passes max recording time the recording will
automatically stop.
-
"No Limit" - Recording will proceed until the internal memory is full. If the recording is
longer than 30 minutes it will be saved in sections, each 30 minutes long.
-
A warning message will be displayed when the available memory is low.
State: Released
Date: 2019.03.06 22:30 GMT