21
incorrect value; see Display Settings (page 44).
Zero the Aortic Pressure Transducer
The
preparation screens
can be used to guide you through the procedure set up. If the guide is
used, follow the instructions on screen. Activating/deactivating the guide is done in the
(Setup)
Menu,
Display
menu,
Other
.
1.
Place the AO transducer at heart level.
2.
Open the AO transducer manifold to air.
3.
Press the
Zero Pa
button. When the text on the
Pa
-button changes to
Zero OK
close the
transducer manifold (open to patient).
Figure 7. Zeroing the Pa
Connect the PressureWire™ Guidewire
1.
Unpack PressureWire™ guidewire using regular sterile routine. Leave the PressureWire
guidewire in the packaging coil. Place the packaging coil flat on the table.
Figure 8. Preparing the PressureWire™ Guidewire
1.
Flush the packaging coil with saline solution through the luer connector.
2.
Continue with either a wireless or cable connected PressureWire guidewire
-
A PressureWire™ wireless guidewire:
Press the
Pd
button in the drop-down menu; press
Connect
(or in the FFR guide press
State: Released
Date: 2019.03.06 22:30 GMT