Table 32. Pressure signal on the HRS via AUX OUT
Erroneous
function
Possible cause
Check
Normal function Corrective action
Bad contact in
cable connectors
Stress the actual
connection
mechanically
The pressure
signal on the HRS
will remain
correct
Replace the
broken cable
AUX OUT cable
not compatible
with HRS
AUX OUT cable
compatibility with
the present HRS
AUX OUT cable is
compatible with
HRS
Replace with a
correct AUX OUT
cable
Defective AUX
OUT cable
Visually check the
monitor cable
Strain relief intact,
no other damage
Replace the AUX
OUT cable
Defective AUX
OUT
Test AUX OUT
according to
Testing AUX OUT
1 & 2 (page 73)
Test of AUX OUT
passed without
failure
Contact St. Jude
Medical
Defective HRS
input channel
Test AUX OUT
according to
Testing AUX OUT
1 & 2 (page 73)
Test of AUX OUT
passed without
failure
Contact
manufacturer of
the HRS
AUX OUT not
configured
correctly
Configuration of
AUX OUT in
Room Setup
AUX OUT
configured
correctly
Edit Room Setup
and correct AUX
OUT
configuration
Table 33. DICOM worklist function
Erroneous
function
Possible cause
Check
Normal function Corrective action
No worklist
received
Network error
Network cables
and components
Cables connected
and network
components
functional
Exchange broken
network cables
and components
Error in network
configuration
Network
configuration in
System menu
N/A
Correct network
configuration in
System menu
DICOM host error Status of DICOM
host
DICOM host is
functional
Correct error in
DICOM host
DICOM worklist
configured
incorrectly
Run verify
function in
DICOM Worklist
setup in System
menu
Verify successful Correct DICOM
worklist
configuration
State: Released