75
Leakage Adapters
Leakage adapters to test patient connections, PW IN/OUT, are included in service kit. The leakage
adapters are made for easy connection between the QUANTIEN™ measurement system and
external equipment during measurement of leakage currents. The input leakage adapters contain
resistors protecting the sensitive components inside the QUANTIEN measurement system during
the leakage measurement.
Service or Repair
Preventative maintenance can be performed by a qualified hospital technician; however, service
or repair of the QUANTIEN™ measurement system must only be performed by personnel
authorized by St. Jude Medical.
Contact St. Jude Medical if the QUANTIEN measurement system requires service, repair or if
there is reason to believe the system's security could be compromised.
Technical Specifications
Pressure Measurement
Table 43. Pressure Measurement
Pressure range
-30 to 300 mmHg
Accuracy:
PressureWire™
guidewire
6
± 1 mmHg plus ± 1% of reading (over the pressure range
-30 to 50 mmHg)
± 3% of reading (over the range 50 to 300 mmHg)
AUX IN 1 (AO)
± 1 mmHg or 0.5%, whichever is greater
Frequency response
0-25 Hz
Resolution
≤ 0.2 mmHg
Electrical Specifications
Table 44. Electrical specifications
Mains classification
Class II Isolation, See Isolation Diagram (page 78)
Operating voltage
100-240 V, 50-60 Hz
Rated current
0.2 A (230 VAC), 0.35 A (110 VAC)
Rated Power Consumption
20 W
Specifications Interfaces
PW IN
Proprietary PressureWire™ interface
Isolation: IEC 60601-1 Type CF Defibrillation protected
PW OUT
According to ANSI/AAMI BP22:1994
Excitation voltage: 2.4 to 8 Volt DC
6
The accuracy is given for the combination of the QUANTIEN™ measurement system and PressureWire™ systems.
State: Released
Date: 2019.03.06 22:30 GMT