Table 19. System on-screen messages
Typ
e
Current Message
Reason
Action
Cancel
OK
E
"Network update failed.
Failed to save network
settings to file system"
Failed to write
network
settings to the
appropriate file.
File is missing
or corrupt.
Contact
St. Jude
Medical
N/A
Confirm and
close window
E
"Network update failed.
Failed to retrieve network
settings."
Failed to read
network
settings from
the appropriate
file. File is
missing or
corrupt.
Contact
St. Jude
Medical
N/A
Confirm and
close window
E
"Network update failed.
Failed to connect to
network with new settings"
Network not
available or
network
settings not
correct.
Check
network
settings
Check
network
connection
N/A
Confirm and
close window
I
"No response on echo
request"
Failed to verify
settings for
DICOM SCU
Check
settings
N/A
Confirm and
close window
I
"No response on echo
request"
Failed to verify
settings for
DICOM export
Check
settings
N/A
Confirm and
close window
I
"Failed to write image to
DICOM peer"
Export to
DICOM archive
failed
Check
settings
N/A
Confirm and
close window
I
"Could not connect to
network path"
Failed to verify
settings for
network path.
Settings
incomplete or
missing
Check
settings for
network
Check
settings for
network path
Check
network
connection
N/A