15
Table 10. Placement Options
Options
Benefits
Limitations
Desktop
−
No equipment on bed rail
−
Easy access to network/video
connectors
Non-sterile nurse can operate
−
Display on cath lab monitor
−
Sterile operator limited to remote
control – limited functionality
−
Limited to wireless PressureWire™
guidewire
Wall
−
No equipment on bed rail or
table surfaces
−
Easy access to network/video
connectors
−
Non-sterile nurse can operate
−
Display on cath lab monitor
−
Sterile operator limited to remote
control – limited functionality
−
Limited to wireless PressureWire
guidewire
Bedside
−
Close to operator
−
Sterile operator can control all
features
−
Allow use of both wireless and
cable connected
PressureWire guidewire
−
Display on cath lab monitor
−
Instrument on bed rail
−
Limited access to network/video
connectors
Mobile
(mounted on
IV-pole)
−
One instrument for multiple
rooms – cost effective
−
Allows use of both wireless
and cable connected
PressureWire guidewire
−
Longer set up time
−
Floor interference
−
Limited access to network/video
connectors
1.
Attach the mounting bracket to the back of the QUANTIEN main unit following the
instructions provided with the bracket.
2.
Mount the QUANTIEN main unit in the desired place in the cath lab. Do not tilt the main unit
backward more than a maximum of 45º from the vertical line. Do not mount the main unit
tilted forward to prevent risk of ingress of liquids, such as IV-infusion etc.
WARNING: Do not use QUANTIEN measurement system if it has been exposed to
liquids when NOT mounted according to above. This can cause the user or patient
to be exposed to electric shock or cause faulty readings of data. Contact
St. Jude Medical representative or technical support for further instructions.
If a table top stand is used, the table should be horizontal. Ensure that the mounting is
secure and that the position of the QUANTIEN main unit is stable.
3.
Connect the mains cable from the mains power connector on the power supply unit to a
mains power outlet. Connect in such way that the mains cable can be easily removed from
the wall power outlet. For identification of correct mains cable, see the table, Identification of
mains cables.
NOTE: The equipment conforms to specifications when operated in a temperature
range of +15°C to +35°C and relative humidity range of 30% to 75% and in an
atmospheric pressure range of 525 mmHg to 795 mmHg.
State: Released
Date: 2019.03.06 22:30 GMT