NOTE: St. Jude Medical has preloaded the QUANTIEN™ measurement system with
the recommended values for the security settings.
Password policy
-
Enforce password history
-
Maximum password age (days)
-
Minimum password length (characters)
-
Password must meet complexity requirements
NOTE: Passwords should be at least 7 characters and contain 3 of the following:
A lowercase character
An uppercase character
A number
A symbol or special character
Account lockout policy
-
Account lockout threshold
-
Account lockout duration (minutes)
Inactivity timeout
-
Timeout period (minutes)
Connections
Installation - Warnings and Precautions
Warnings
External equipment intended for connection to the QUANTIEN™ measurement system shall
comply with relevant IEC standard (e.g., IEC 60601 series for medical electrical equipment).
In addition, all such combinations of systems shall comply with the standard IEC 60601-1-1,
Safety requirements for medical electrical systems, alternatively IEC 60601-1 ed.3 §16, ME
SYSTEMS. Any person who connects external equipment to signal output, or other
connectors, has formed a system and is therefore responsible for compliance of the system
with these requirements. If in doubt, contact a qualified technician, or alternatively, a
St. Jude Medical representative/technical support.
No modification of this equipment is allowed.
Do not open or remove access covers on the QUANTIEN measurement system unless
specifically instructed by St. Jude Medical technical support to do so.
The AUX inputs/outputs are isolated with one means of operator protection (MOOP); see the
Isolation Diagram (page 78). External equipment which is connected to these inputs/outputs
must provide isolation against leakage current. The user is responsible for compliance with
the requirements of standard IEC 60601-1-1, Safety requirements for medical electrical
systems, alternatively IEC 60601-1 ed.3 §16, ME SYSTEMS.
Cautions
Do not connect patient applied parts to AUX IN/OUT analog ports.
Output sensitivity changes should be performed by a qualified technician. Incorrect settings
may result in discrepancies between monitor system and QUANTIEN™ measurement
State: Released
Date: 2019.03.06 22:30 GMT