Add/Delete Room
At initial setup:
The window will be empty if no Wi-Box™ AO transmitter is within radio range. Touch +
Add
to add and configure a new room.
If one or more Wi-Box AO transmitters are turned on and within radio range, the
QUANTIEN™ measurement system will display a room button for each Wi-Box AO
transmitter with the identification number of the Wi-Box device. Push
Edit
and then the
Room
button to configure one room at a time.
If no powered Wi-Box AO transmitter is present or within radio range, a new room will have
to be setup manually.
Figure 19. Room Button Menu
1.
Name for Room
2.
Use as Default
Add/Edit Room:
Touch
Name
field and type a room name (maximum 8 characters).
Confirm that the Wi-Box AO transmitter identification number matches the number printed
on the Wi-Box unit in the current room.
Touch the
AO
button to select cabled analog input AO source. Refer to Connect AO Via
Analog Input AUX IN 1 (page 38), for configuration of analog input as source for AO
pressure.
Optional:
Select the
AUX
button to configure analog input/output channels:
In 2
Import ECG signal (see Connect ECG Via Analog Input AUX IN 2 (page 40)).
Out 1
and
Out 2
Export Pd, Pa or Reference signal (see Configure Analog Output Ports AUX
OUT 1 & 2 (page 40)).
Select
Use as Default
to make a room setup default at system startup. When one room is
set as default the
Select Room
window will be bypassed at start up and the
Patient
window
will be displayed directly. This setting is preferred when the QUANTIEN measurement
system is permanently installed in a room.