The following settings are required:
Date limitation options
-
Procedures Scheduled for Today
-
Procedures in the Range:
-
Enter number of Days Before Today´s Date
-
Enter number of Days After Today´s Date
-
Do Not Consider Date
Scheduling options
-
Procedures Schedule for Modality (the Modality selected earlier is displayed for
information)
-
Procedures Scheduled for AE Title (the AE Title entered earlier is displayed for
information)
-
Procedures Scheduled for Station Name (the Station Name entered earlier is displayed
for information)
Maximum Retrieved Procedures
-
Enter value to set maximum number of retrieved procedures.
Configure Export to DICOM Archive
Enter
(Setup)
Menu,
System, DICOM, Export
Settings made here configures export to a DICOM Archive.
The following information is required:
Local AE Title
Remote AE Title
Remote Host (Hostname/IP Address)
Remote Port Number
Modality
: Select either XA or OT Modality
Press the
Verify
button to apply settings and verify that the connection works.
Institution Name:
If required a specific Institution name can be configured in
(Setup)
Menu,
System, Info,
Institution
. The Institution name will be included in the file information when exporting to
DICOM.
Display Settings
Unless otherwise indicated, settings are entered in
(Setup)
Menu,
Display
menu.
Sweep Speed
Change screen update rate and the level of detail visible to the user.
A high sweep speed is suitable when a detailed picture of the tracings is required.
A low sweep speed is suitable when displaying slow changes, for instance during intravenous
infusion and pressure pullback.
NOTE:
Due to the lower update rate at slow and medium sweep speed, some information
may not be fully displayed.
State: Released
Date: 2019.03.06 22:30 GMT