8
Table 3. Devices and Accessories Manufactured by St. Jude Medical
Device
Comments
Mounting bracket
Required. Several models available. Wall mount, desktop stand,
bedside pole bracket and pole bracket.
PressureWire™ guidewire Required. Use the PressureWire™ Aeris™ guidewire or
PressureWire™ X wireless guidewire for wireless connection. Use the
PressureWire™ Certus™ guidewire or PressureWire™ X cable
connection guidewire for connection by cable.
Wi-Box™ AO transmitter
and cables
Required for wireless configuration. Interface for wireless transmission
of AO pressure to the QUANTIEN™ measurement system. Adapter
and monitor cables are required for connection to HRS.
RadiView™ software
Optional. Software for installation on a PC. Allows for post-procedural
reviewing and analysis of FFR measurements.
AUX IN/ECG cables
Optional. For transmission of AO pressure and ECG signals to the
QUANTIEN measurement system from the HRS, specific cables for
each system.
PressureWire Out cable
Optional. For transmission of distal pressure from the QUANTIEN
measurement system to the HRS, cables specific for each system.
Service kit
Used when testing leakage current and PW IN/OUT on the
QUANTIEN measurement system. Also used for testing AO IN/OUT on
the Wi-Box AO™ transmitter.
AUX Out cable
Optional. For export of Pa or Pd analog signal to an external
measuring instrument.
Table 4. Devices and Accessories from Other Suppliers
Device
Comments
Aortic pressure (AO)
transducer
Required. Transducer shall comply with the standard for invasive
blood pressure transducers, ANSI/AAMI BP22-1994.
DVI-I cable
Optional. Video cable to display QUANTIEN™ screen on a cath lab
monitor or external video monitor (e.g., conference room).
Ethernet cable
Optional. Network cable. QUANTIEN measurement system network
connection (RJ45).
Mounting bracket
If other than St. Jude Medical standard bracket is used, it must
comply with VESA standard 75x75 mm or 100x100 mm.
USB memory stick
Only a port-powered USB memory stick is allowed.
General Warnings, Precautions, and Safety Information
Warnings
No modification of this equipment is allowed.
Do not open or remove access covers on the QUANTIEN™ measurement system unless
specifically instructed by St. Jude Medical technical support to do so.
External equipment intended for connection to signal input, signal output or other
State: Released
Date: 2019.03.06 22:30 GMT