Table 9. Identification of Mains Cables
Australia
Rated Voltage: 250 V
Rated Current: 2.5 A
NOTE: If a cable needs to be replaced the cord set shall fulfill applicable regulations. If
in doubt please contact St. Jude Medical.
CAUTION: For markets other than those specified in the table above, make sure
the chosen cord set fulfills applicable regulations.
To be able to mount and install the QUANTIEN main unit, a mounting bracket is required. Use
brackets supplied separately from St. Jude Medical or use an existing bracket compatible with
standard VESA 75 x 75 mm or 100 x 100 mm. Refer to the Placement Options table for possible
mounting solutions.
Before proceeding with installation, check that all parts are undamaged.
Mount QUANTIEN™ Main Unit in the Cath Lab
WARNING:
Make sure the equipment is securely mounted. If the equipment is not
securely attached, it may fall over causing possible patient or operator injury
and damage to the system.
Do not use the QUANTIEN™ measurement system if it has been dropped or
in any other way exposed to mechanical or electrical damage. This can cause
the user or patient to be exposed to electric shock or cause faulty readings of
data. Contact St. Jude Medical representatives or technical support for
further instructions.
Do not use the QUANTIEN measurement system if there is reason to believe
the system's security has been compromised or if the system was
unaccounted for a period of time (i.e. misappropriated, modified or tampered
with).
To protect the privacy and security of sensitive information, including
electronic protected health information (EPHI), and to protect the integrity of
the system itself, the system should be located in a physically secure, access
controlled environment.
CAUTION: Make sure a power cable with a connector suitable for the mains power
outlet is used. Verify that the chosen cord set fulfills applicable ratings and
regulations.
St. Jude Medical recommends the QUANTIEN measurement system is installed in a secure,
access controlled environment. If the QUANTIEN measurement system is not permanently
mounted, it should be stored in a secured, access controlled location.
The design of the QUANTIEN measurement system allows flexible placement either inside or
outside the patient environment with fast and easy setup regardless of position. The placement in
the cath lab will impact which devices, wireless or wired, can be connected. Benefits and
limitations for different placement are summarized in the table below.