Table 22. PW Source — PressureWire™ Wireless Guidewires
Type
Current Message
(on Pd button)
Event
Action
W
(Yellow frame)
(Pd Not Zeroed)
Automatic zeroing after
successful mate failed
Ensure the
PressureWire™
guidewire is correctly
prepared
Try to connect again
W
(Yellow frame)
The PressureWire
transmitter is running
low on battery
Replace the
PressureWire
guidewire
E
(Red frame)
(Pd Connect failed)
The system cannot
connect to the
PressureWire
guidewire
Ensure that the
PressureWire
transmitter is turned
on at the appropriate
time
Ensure that it is not too
far away from the
system
E
(Red frame)
(Pd Signal Lost)
Communication with
the PressureWire
guidewire
broken/interrupted
The PressureWire
wireless guidewire
turned off
Turn the PressureWire
guidewire on
Remove reason for
interruption/breakage
of communication
E
(Red frame)
(Pd Interface Error)
The PressureWire
interface missing
The PressureWire
interface reports error
Restart System
Contact St. Jude
Medical
E
(Red frame)
(Pd Zero Failed)
The PressureWire
guidewire not
positioned correctly
(e.g., inserted in the
blood flow) when
zeroing
Ensure the
PressureWire
guidewire is correctly
prepared and
positioned
Try to zero again
Table 23. PW Source — PressureWire™ Cable-Connection Guidewire
Type
Current Message
(on Pd button)
Event
Action
State: Released
Date: 2019.03.06 22:30 GMT