Table 27. Pressure signal on QUANTIEN™ measurement system when acquired via AUX IN
function
Erroneous
function
Possible cause
Check
Normal function Corrective action
Unable to zero
AUX AO source
Varying pressure
signal output from
HRS
Refer to IFU for
HRS on how to
check pressure
output signal
Stable pressure
signal
Refer to IFU for
the HRS on how
to correct
pressure output
signal
Too high or too
low pressure
signal output from
HRS
Refer to IFU for
HRS on how to
check pressure
output signal
Pressure signal
close to zero
Refer to IFU for
the HRS on how
to correct
pressure output
signal
Unable to
calibrate AUX AO
source
Varying pressure
signal output from
HRS
Refer to IFU for
HRS on how to
check pressure
output signal
Stable pressure
signal
Refer to IFU for
the HRS on how
to correct
pressure output
signal
Too high or too
low pressure
signal output from
HRS
Refer to IFU for
HRS on how to
check pressure
output signal
Pressure signal
close to selected
calibration level
Refer to IFU for
the HRS on how
to correct
pressure output
signal
Mismatch
between selected
calibration level
and applied
pressure signal
from HRS
Compare selected
calibration level
and applied
pressure signal
from HRS
Pressure signal
close to selected
calibration level
Select correct
calibration level
Unable to zero Pa
pressure
Varying pressure Check that AO
transducer
manifold is
opened to air
AO transducer
manifold is
opened to air
Open AO
transducer
manifold to air
Too high or too
low pressure from
AO transducer
Check that AO
transducer is
placed on correct
level
AO transducer is
placed leveled
with patient
Reposition AO
transducer
Disturbance to Pa
pressure signal
Check Pa
pressure signal
for variations
No variation in Pa
pressure signal
Check according
to "Abnormal or
absent pressure
reading of Pa
pressure"
Abnormal or
absent pressure
reading of Pa
pressure
Bad contact in
AUX IN cable
connectors
Stress the actual
connection
mechanically
The Pa pressure
signal will remain
correct
Replace the
broken cable
Defective AUX IN
cable
Visually check the
cable
Strain relief intact,
no other damage
Replace the
broken cable
State: Released