Table 35. External display function
Erroneous
function
Possible cause
Check
Normal function Corrective action
Defective external
monitor
Test with another
external monitor
QUANTIEN
display viewed on
external monitor
Replace the
external monitor
External monitor
not compatible
with DVI-I signal
from the
QUANTIEN
measurement
system
Monitor input
supports signal
properties
according to
Specifications
Interfaces (page
75).
N/A
Replace the
external monitor
Table 36. Other functions
Erroneous
function
Possible cause
Check
Normal function Corrective action
Equalize fail
Signal error or not
zeroed Pd or Pa
Status of Pa and
Pd
Pa and Pd zeroed
and functional
Correct errors
and/or perform
zeroing of Pa and
Pd
Too large
difference
between Pa and
Pd
Difference
between average
of Pa and Pd
Difference
between average
of Pa and Pd is
less than
30 mmHg
Reposition AO
transducer and
perform a re-
zeroing of Pa
Re-zeroing of Pd
has been
performed when
using the wireless
PressureWire™
guidewire
N/A
N/A
Re-initialize the
wireless
PressureWire
guidewire by
turning it off and
then on again
Recording ends
without being
stopped by user
Recording time
limited in setup
Max recording
length in System
menu (Other)
N/A
Set appropriate
max recording
length
Memory full
Storage space in
Archive
Storage space is
available
Free-up storage
space
Maintenance
Remote Control Battery
Tools needed:
Phillips screwdriver
New batteries, 2 x AAA (LR03, 1.5V)
State: Released
Date: 2019.03.06 22:30 GMT