Overview:
Connecting the QUANTIEN measurement system to a network will allow data export for remote
review or archiving, communication with a DICOM server for import of a DICOM worklist and
screen image export to DICOM compatible archives.
Connection:
1.
Connect the ethernet patch cable with isolation interface, provided with the QUANTIEN
measurement system, to the ethernet port (Product Interface and Symbols (page 3)).
2.
Connect a standard network (RJ45) cable between a hospital network port and the isolation
interface.
Configuration:
Setup, System, Network
Required information:
Hostname:
Manually enter a local name for QUANTIEN measurement system
IP Address:
Option to enter manually or to obtain automatically
Subnet Mask
Gateway
Primary DNS Server
Press
Apply
to apply network settings.
Configure Data Export
Enter
(Setup)
Menu
, System, Export
Network Path
Configure the network path to which the QUANTIEN™ measurement system will export data,
either manually or automatically. Requires a functional network connection.
Annotation: //"server name"/"folder X"/"folder Y". If backslash (\) is used instead of slash (/), the
system will automatically change them.
Export path:
Network pathway
User name:
Network user name with read/write access rights
Password:
Network user password
Press the
Verify
button to verify that the connection is working.
Export to Compatibility Mode
Data can be exported in a format that is compatible with RadiView™ software version 2.2 and
earlier. For versions later than 2.2, the checkbox can be deselected to allow more information to
be displayed in RadiView software.
Auto Export
Activate/de-activate automatic data export. Select to which destination and in which format the
QUANTIEN™ measurement system will save recordings as soon as they are completed after
review.
Destinations:
Network Path:
Export data to the configured network path
DICOM Archive
: Export data to the configured DICOM Archive
State: Released
Date: 2019.03.06 22:30 GMT