51
Table 20. AO Source — Wi-Box™ AO Transmitter
Type
Current Message
(on Pa button)
Event
Action
E
(Red frame)
(Signal Lost)
No communication with
the Wi-Box™ AO
transmitter
Turn Wi-Box unit on
Remove reason for
interruption/breakage of
communication
Select a room with a working
Wi-Box unit
E
(Red frame)
(Wi-Box Error)
Faulty signal/data from
Wi-Box AO transmitter
Restart Wi-Box unit
Restart System
Contact St. Jude Medical
E
(Red frame)
(Pa Interface Error)
AO USB missing
AO USB reports error
Restart System
Contact St. Jude Medical
E
(Red frame)
(Pa Zero failed)
Pressure transducer not
open to air when zeroing
Zeroing offset out of
range
Prepare for zeroing
Zero
Table 21. AO Source — AUX IN
Type
Current Message
(on Pa button)
Event
Action
E
(Red frame)
(Pa Interface Error)
AUX missing
AUX reports error
Restart System
Contact St. Jude
Medical
E
(Red frame)
(Pa Zero Failed)
Pressure transducer
not open to air when
zeroing
Zeroing offset out of
range
Prepare for zeroing
Zero
Table 22. PW Source — PressureWire™ Wireless Guidewires
Type
Current Message
(on Pd button)
Event
Action
State: Released