73
Testing AUX IN 2 and Associated Monitor Channel
Tools needed
AUX IN 2 cable
HRS
Service report form, can be downloaded from sjmprofessional.com
Preparation
Switch on QUANTIEN™ measurement system at least 10 minutes before starting this test.
Connect the AUX IN 2 cable from the HRS to AUX IN 2.
To test the AUX IN 2, proceed as follows:
1.
Restart QUANTIEN measurement system by switching the main switch OFF. Wait for 5
seconds and switch ON again. Verify that passed self-test is indicated.
2.
Select the applicable room and check that the room is configured for use of ECG via AUX
IN2.
3.
Select patient, e.g., default patient.
4.
Apply an ECG signal from the HRS, e.g., by using a patient simulator with the HRS.
5.
Check that the signal displayed in QUANTIEN measurement system corresponds with the
signal in the HRS. Possibly adjust the ECG display scale to make an assessment of the
amplitude. Note the result of the test in the service report.
Testing AUX OUT 1 & 2 and Associated Monitor Channel
Tools needed
AUX OUT cable
A measuring device, like a multimeter.
Service report form, can be downloaded from sjmprofessional.com
Preparation
Switch on QUANTIEN™ measurement system at least 10 minutes before starting this test.
Connect the AUX OUT cable from AUX OUT 1 to a measuring device.
To test the AUX OUT, proceed as follows:
1.
Restart QUANTIEN measurement system by switching the main switch OFF. Wait for 5
seconds and switch ON again. Verify that passed self-test is indicated.
2.
Select to edit the applicable room.
3.
Select "AUX".
4.
Select "OUT 1".
5.
Select "Reference".
6.
Set 0.0 V as the output signal on QUANTIEN measurement system.
7.
Read the value on the measuring device. If necessary convert the reading value to the
corresponding input voltage. Record the converted value in the service report and check that
it is between –50mV and +50mV.
8.
Set 1.0 V as the output signal on QUANTIEN measurement system.
9.
Read the value on the measuring device. If necessary convert the reading value to the
State: Released
Date: 2019.03.06 22:30 GMT