16
Powering On and Shutting Down
Powering On
Check that the mains cable is connected to the power supply unit and to the mains power outlet.
Press the power switch on the bottom panel to start the QUANTIEN™ measurement system.
The screen will remain dark for a few seconds before a startup beep indicates that the system is
booting up.
When the status check is OK, indicated by blue checkmarks, the
User Login
window appears.
Enter your credentials to log into the system. See User Settings for more information (page 35).
WARNING: Do not leave the QUANTIEN measurement system unattended while
logged in as a Site Administrator.
NOTE: If you do not have credentials to log into the system, you can log in as "guest".
After login, the
Select Room
window appears. If a room configuration is set as default then the
first screen will be the
Patient
menu. Select
Skip
to move to the
(Setup)
menu where any
menu can be selected.
At interruption of supply mains for less than 5 minutes an ongoing examination can be continued
seamlessly provided that both Pa and Pd have been zeroed before power was lost. After power
recovery you will be asked if the previous examination shall be continued. If you select to
continue, the examination will continue. Otherwise the system will perform a normal start up. If
supply mains is interrupted for more than 5 minutes, the QUANTIEN measurement system will
always perform a normal start up.
Shutting Down
Press the switch on the bottom panel to OFF to shut down.
CAUTION: The power switch will shut down the QUANTIEN™ measurement
system but there is still power in the mains cable and power supply unit. To fully
disconnect from mains power, remove the mains power plug from the wall socket.
Room Configuration
After logging into the QUANTIEN™ measurement system, the
Select Room
window should open
(alternatively select the
(Setup)
Menu, choose
Select Room
to open the
Room
window).
In the
Select Room
window a button for each transmitting Wi-Box™ AO transmitter within radio
range of the QUANTIEN measurement system is displayed. On each button the identification
number of the transmitting Wi-Box AO transmitter is displayed.
NOTE: All Wi-Box AO transmitters transmitting in the vicinity of the QUANTIEN
measurement system will be displayed. Check that the correct Wi-Box unit is linked to
the correct room when you configure each room; each Wi-Box unit has the
corresponding identification number printed on the front.
1.
To link a Wi-Box AO transmitter to a room, press first the
Edit
button, and then the
Room
button with the identification number of the correct Wi-Box AO transmitter.
NOTE: The
Edit
button is not available to all users. Refer to User Settings for more
information. (page 35)
2.
Press the
Name
field and enter the
Room
name (maximum 8 characters).
3.
Select
Use as default
if the QUANTIEN measurement system is placed permanently and
this room should automatically be selected at startup.