Table 25. Remote control function
Erroneous
function
Possible cause
Check
Normal
function
Corrective action
No or intermittent
functioning of the
remote control
Low battery
N/A
N/A
Replace the
batteries, see
Remote Control
Battery (page 67)
Defect remote
control unit
Check functioning
with another
remote control
unit
The system
responds
normally
Replace the
remote control unit
Internal failure in
main unit
N/A
N/A
Contact St. Jude
Medical
Ambient light
interference
Switch off
ambient light.
Check functioning
The system
responds
normally
Decrease ambient
light intensity or
switch to another
ambient light
source
Table 26. Pressure signal on QUANTIEN™ measurement system when acquired via Wi-Box AO
transmitter
Erroneous
function
Possible cause
Check
Normal function Corrective action
Wi-Box™ AO
transmitter not
possible to select
for a room
Wi-Box unit not
powered on or
internal failure in
Wi-Box unit
Check LED’s of
Wi-Box AO
transmitter
Refer to IFU for
the Wi-Box AO
transmitter
Refer to IFU for
the Wi-Box AO
transmitter
Internal failure of
AO part in
QUANTIEN™
main unit
Check HW status
of AO after-self
test at power on
of QUANTIEN
main unit
AO status OK
Contact St. Jude
Medical
Unable to zero Pa
pressure
Varying pressure Check that AO
transducer
manifold is
opened to air
AO transducer
manifold is
opened to air
Open AO
transducer
manifold to air
Too high or too
low pressure from
AO transducer
Check that AO
transducer is
placed on correct
level
AO transducer is
placed level with
patient
Reposition AO
transducer
Disturbance to Pa
pressure signal
Check Pa
pressure signal
for variations
No variation in Pa
pressure signal.
Check according
to "Abnormal or
absent pressure
reading of Pa
pressure"