Table 11. Explanation of Live Window Interface
11
Main function button
NOTE: Appearance/function changes depending on current mode.
Zero Pa
to zero the AO transducer during preparation step 1.
Connect
to connect the PressureWire™ guidewire during preparation step 2.
Equalize
to equalize the Pd and Pa during preparation step 3.
Record
to start recording during an FFR procedure
Pullback
to start recording during a RFR procedure
Stop
to end a recording of an FFR or RFR procedure
Cancel
to end a snapshot recordings
Live
to return to the live window during review or settings menus
12
NOTE: Appearance/function changes depending on current mode.
Prep
button – returns to the preparation procedure while there is no current recording
Mark
button – Inserts marker in the recording/pullback only.
13
Vessel selection menu - touch for drop-down menu.
14
Menu bar
– move to other menus/windows.
NOTE:
The menu bar is grayed out during procedure recordings.
Some buttons may not be available for all users. Refer to User Settings for
more information. (page 35)
Depending on the step in the procedure additional buttons/fields may appear
AO source not available or not selected (red frame).
Time equalize function failed or is turned off and may affect the accuracy of the
calculated index (yellow)
Color Coding
Red represents the Pa measurement on the screen. However, a red frame around any of the
boxes on the right indicates an error. See Troubleshooting for more information (page 47).
Green lines on the graph represent Pd measurements.
Yellow lines on the graph represent Pd/Pa or FFR measurements. However, a yellow frame
around any of the boxes on the right indicates an error. See Troubleshooting for more information
(page 47).
Blue lines on the graph represent RFR measurements.
During setup the Pd and Pa buttons display status information, the colors of the frame indicate
status changes. Red frame indicates connection/signal lost, yellow frame indicates that setup is
needed and green frame flashes during connection/zeroing etc.
FFR/RFR Measurement Procedure
NOTE: Reference the IFU for the PressureWire™ guidewire for complete instructions
State: Released