41
Connect Video Out to External Monitor
WARNING: The DVI-I output is not isolated. External equipment which is
connected to this output must provide isolation against leakage current. The user
is responsible for compliance with the requirements of standard IEC 60601-1-1,
Safety requirements for medical electrical systems, alternatively IEC 60601-1 ed.3
§16, ME SYSTEMS.
The DVI-I (Integrated) video output port enables the QUANTIEN™ screen to be displayed on an
existing cath lab monitor or external slave monitors.
Screen resolution of QUANTIEN main unit: 1280x800.
Connection:
Connect the video port on the receiving monitor/equipment to the DVI-I port (Product Interface
and Symbols (page 3)) on the bottom panel of the QUANTIEN main unit using a standard video
cable.
NOTE: The QUANTIEN measurement system will automatically detect the external
monitor's resolution.
Export Recordings to USB Memory Stick
Connect a regular USB memory stick to the USB port (Product Interface and Symbols (page
3)).
Use the
Export
button in either the
Review
or
Archive
window to export data to the USB
device, see Export Data (page 28).
WARNING: Any USB memory stick used must be port powered if connected to the
QUANTIEN™ measurement system in the patient´s vicinity or it may compromise
electrical isolation and cause patient injury. It is not allowed to electrically connect
the USB device to other devices.
Network Configuration
Configure General Network Settings
WARNING: The ethernet patch cable with isolation box must be used to assure
proper isolation as the ethernet outlet is not isolated. The short patch cable shall
be positioned between the isolation box and the main unit. The user is responsible
for compliance with the requirements of standard IEC 60601-1-1, Safety
requirements for medical electrical systems, alternatively IEC 60601-1 ed.3 §16,
ME SYSTEMS.
CAUTION: Direct connection to a non-secure network, like the internet, may
interfere with correct operation and/or result in inappropriate access to patient
information. Furthermore, it should be noted that reconfiguring a used network
may lead to inability to import patient as well as export examination data,
ultimately leading to a risk of loss of patient and examination data. To avoid this
problem St. Jude Medical recommends verifying network settings in the system
setup after each change.
NOTE: Please note that St. Jude Medical makes no representation of warranty that use
of the QUANTIEN™ measurement system complies with applicable privacy, security
and confidentiality laws, but encourages you to assess your own risk as you use,
disclose, control, process or transfer patient health information with this system.
State: Released
Date: 2019.03.06 22:30 GMT