74
corresponding input voltage. Record the converted value in the service report.
Check the value, subtracted with the value read for 0.0 V reference signal, is between 0.95
and 1.05 V.
10.
Set "Off" as the output reference signal on QUANTIEN measurement system.
11.
Connect the AUX OUT cable from AUX OUT 2 to a measuring device.
12.
Select "OUT 1".
13.
Select "AUX".
14.
Select "OUT 2".
15.
Repeat steps 5 to 9.
NOTE: Limits for readings do not include tolerances for possible deviations in the
measuring device. Therefore, if any failing measurement is observed, the measuring
device should be checked.
Test Probes
Introduction
The PW control probe is a tool for accuracy control of the PressureWire™ input on the
QUANTIEN™ main unit. The green PW control probe simulates a PressureWire™ sensor.
Environmental Conditions
Table 41. Operating Conditions
Operating, temperature:
15°C to 30°C (59°F to 86°F) max 1°C temperature change
during calibration control.
Operating, relative humidity:
30-75%
Transport and storage, temperature: -40°C to + 70°C
Transport and storage, relative
humidity:
10% to 100%
Cleaning:
Clean the control probes with a moist tissue, using a mild
detergent. Do not use alcohol-based solutions.
Table 42. Specifications PW Control Probe
Pressure levels
-25 mmHg (Black button)
0 mmHg
(No button)
295 mmHg
(Red button)
Accuracy:
<0.33 mmHg
When using the control probe for accuracy control, the probe must be at the normal temperature
for at least 30 minutes before use.
The PW control probe has been designed with only passive components and molded plastic in
order to achieve maximum robustness. Therefore, the control probe does not need to be
calibrated. If any control point measurement fails during regular maintenance, please return the
QUANTIEN measurement system together with the control probe used.
State: Released
Date: 2019.03.06 22:30 GMT