20
on the FFR/RFR procedure.
Overview of a measurement procedure:
1.
Start system, log into system and select room
2.
Select or create new patient
3.
Zero the aortic pressure transducer
4.
Connect the PressureWire guidewire
5.
Equalize Pa and Pd pressures
6.
Select recording mode
7.
Perform an FFR recording or an RFR pullback
8.
Review and annotate measurement recording
9.
Export Data
Start System and Select Room
Preparations
Observe normal sterile procedures if the equipment is used within the sterile zone. This includes
covering the QUANTIEN™ measurement system in sterile plastic and keeping the remote control
in a sterile plastic bag. Please refer to Cleaning (page 68) for further instructions on cleaning of
the QUANTIEN measurement system.
Start System and Select Room
Ensure that the QUANTIEN™ measurement system is connected to mains socket and turn on the
device, the
Select Room
screen will be displayed after logging into the system.
Select the room to be used by touching the corresponding
Room
button.
NOTE: If the room has been set as default this screen will not be displayed.
For detailed instructions, refer to Room Setup (page 32).
Select or Create New Patient
After room selection the
Patient
window is displayed. There are three optional ways to create or
select a patient for the new study:
1.
Create new patient: Touch +
New Patient
and enter the patient information manually.
NOTE: You must create a new patient when logged in as a Guest. See User Settings for
more information (page 35).
2.
Create
New Study
for existing patient: Open the
Archive
window, choose an existing patient
entry and choose
New Study
.
3.
Select patient from DICOM Worklist: Select a patient entry and choose
New Study
.
For detailed information on DICOM settings and worklist query, see Network Configuration (page
41).
After patient selection the
Live
window will be displayed.
Verify measurement settings, see Display Settings (page 44).
CAUTION: An insensitive or overly sensitive averaging of pressure may result in an
State: Released
Date: 2019.03.06 22:30 GMT